|
Specifications
Operation
Troubleshooting
Specifications:
Does the
Pressure Injection Cell require electrical power?
No. The liquid sample
is forced through the capillary using pressurized gas
typically supplied by a tank.
Back to Top
Operation:
How much
pressure is required?
Loading samples into a
capillary for mass spectroscopy typically requires 100 to
400 psi. Packing capillary columns typically
requires 500 to 1000 psi.
How long
does packing a column take?
Packing a 20 cm column
typically takes a few hours. When you are finished
packing the column, turn off the pressure at the gas tank
and let the column sit overnight so that the pressure in
the cell leaks out through the packing. This avoids
pull back of the packing media.
How can you
determine the approximate flow rate through a capillary?
If the solution is flowing
only through an otherwise empty capillary tube, the flow
rate is straightforward to calculate. However, a
frit or a packed capillary typically causes much more flow
resistance, so it is best to measure the flow rate with a
test solution.
To measure the flow rate,
run a test solution through a sample capillary and at the
outlet place a calibrated micropipette. The solution
should wick up the micropipette on its own so you don't
need any special fittings to seal them together. For
example, Drummond and other brands of micropipettes,
available from most laboratory product suppliers, come in
1, 2, 3, 4, 5, 10, 20, ... µl volumes. Using a
stopwatch, measure how long it takes from when the
solution starts to exit the capillary and begins filling
the micropipette until the micropipette is filled.
Back to Top
Troubleshooting
Gas leaks through the top-to-base interface of the pressure
injection cell. The three
bolts might not be fully tightened or the O-ring might be
nicked or dirty or missing. If the O-ring is dirty, wash
it off with a soap and water and return it to its groove.
Contact us for a replacement O-ring.
Gas leaks around the capillary.
The ferrule should be replaced.
You do not see sample traveling through the capillary
after a few minutes.
Check that the gas tank is supplying the desired pressure
(typically on the order of several hundred psi) and that
the 3-way valve knob is pointing toward the pressurized
gas line.
The media
pulls back when you release the pressure.
When you are finished
packing the column, turn off the pressure at the gas tank
and let the column sit overnight so that the pressure in
the cell leaks out through the packing. This avoids
pull back of the packing media.
Back
to Top
|