Next Advance Inc produces: the Freedom Rocker which automatically processes western, southern, and northern blots and gels; laboratory platform rockers; pressure injection cells; and more.

We specialize in Homogenizers, Rockers, Pumps, and Automated Blot Processors.  Call 1.518.674.3510

Automated
Blot Processor
Capillary Loader Syringe Pumps Homogenizer
Cell Disrupter

Innovative Lab Products for the Life Sciences

The Bullet Blender™ and Bullet Blender™ Blue

  

  Want to beat your Fat away?

(don't we all???)

Your Fat/Adipose samples stay cool
during homogenization!



The Bullet Blender™Lyse E. coli and other cells. Homogenize or disrupt tissue. The Bullet Blender is quiet and user friendly.

>
No Cross Contamination

> Samples stay cool
> Convenient
> Risk  Free
> Established protocols

Homogenize up to 24 Fat/Adipose samples simultaneously-- in less than 3 minutes!

 

The Bullet Blender™ enables you to homogenize up to 24 samples in separate microcentrifuge tubes at a time.  Load up to 300mg Fat/Adipose with your application buffer and grinding beads in standard polypropylene microcentrifuge tubes.  Place the tubes into the Bullet Blender™ (pictured above) or Bullet Blender™ Blue The "bullets" in the "blender" vigorously strike all the sample tubes as the instrument runs for a few minutes.  The agitation of the grinding agents with the tissues homogenizes the samples.


The Bullet Blender™ and Bullet Blender™ Blue enable you to disrupt or homogenize up to 24 adipose and other tissue/cell samples at a time.  Load the samples in standard polypropylene microcentrifuge tubes into the Bullet Blender ™.  Balls (the "bullets") repeatedly strike the sample tubes in a controlled manner, thousands of times per minutes, inducing vigorous motion of the tissue samples or other substances inside the tubes, providing efficient mixing. With beads in the tubes, the samples are thoroughly homogenized. Tune the extent of disruption by adjusting the speed. Click here to see sample protocols.


> No Cross Contamination

The samples are homogenized (or mixed) non-invasively.  The sample tubes are kept closed during agitation, as the samples are processed.  There are no probes to clean between samples.

> Samples Stay Cool

The instrument uses very little power due to the unique, patented design.  It uses a small DC motor to agitate the individual tubes, not ultrasonics.   Also, because the motor does not need to agitate any heavy platforms or plates, it is small and will last for years.

> Convenient to Use

Simply place your adipose and other tissue samples or cell cultures and some beads in standard high quality polypropylene tubes, and load the tubes into the Bullet Blender™.   Set the duration (typically a few minutes) and speed (vigorousness).  There are no probes to clean.  And it is quieter than a sonicator and does not heat up your samples more than a few degrees!

> Risk Free Purchase

The Bullet Blender™ comes with a 30-day money back guarantee and a two year warranty, with a three year warranty on the motor.  The simple, reliable design enables the Bullet Blenders™ to sell for a fraction  of the price of ultrasonic or other agitation based instruments, yet provides an easier, quicker technique.

> Established Protocols


Protocol for Adipose Tissue Homogenization in the Bullet Blender™

The protocol described in this document is for the use of the Bullet Blender™ for the homogenization of fat / adipose tissue (from a variety of animals). Note that the time and speed settings may differ due to the variation in consistency/texture of fatty tissue from species to species. This protocol does not specify a particular buffer - you may choose which is most appropriate for your downstream application (nucleic acid isolation, protein extraction, etc.).

Materials Required:

adipose tissue, saline, aspirator, Bullet Blender™, zirconium oxide beads (0.5mm OR 1.0mm) , microcentrifuge tubes, homogenization buffer, and pipetor.

Instructions:
  1. Cut adipose tissue into appropriately sized pieces for analysis (30mg-300mg) and
  2. place into a microcentrifuge tube.
  3. OPTIONAL: Wash tissue with ~1mL PBS. Aspirate. NOTE: This step removes external contaminants (blood, etc.).
  4. Add a mass of 0.5mm zirconium oxide beads equal to the mass of sample in each tube (so for 100mg sample, add 100mg beads). One scoop ≈ 180mg.
  5. Add 2 - 4 volumes of buffer for every volume of tissue.
  6. Close the microcentrifuge tubes.
  7. Place tubes into the Bullet Blender™.
  8. Set controls for SPEED 7 and TIME 2 minutes. Press Start.
  9. After the run, remove tubes from the instrument.
  10. Inspect samples. Fatty tissue homogenate will be difficult to see through due to the light scattering of lipid micelles formed, so it may be necessary to employ a pipette tip to check inside the tube for remaining pieces of intact tissue. If homogenization is unsatisfactory, run for another minute at SPEED 9.
  11. Proceed with your downstream application.
SAFETY NOTE!!!

When using a centrifuge to separate your homogenate from the debris and beads, make sure your tubes are balanced.

Special thanks to Dr. Timothy Bartness and Yang Liu at Georgia State University for their feedback on this protocol.




Protocol for Adipose in Bullet Blender™ 50 Homogenization

The protocol described in this document is for the use of the Bullet Blender™ 50 for the homogenization of Fat / Adipose Tissue. This protocol does not specify a particular buffer - you may choose which is most appropriate for your downstream application (nucleic acid isolation, protein extraction, etc.).

Materials Required:

lung , Bullet Blender™ 50, homogenization buffer, pipettor, 50mL centrifuge tubes, 3.2mm stainless steel beads (part number SSB32).

Instructions:
  1. Cut adipose into appropriately sized pieces for analysis (0.1g – 3g) and place into a 50mL centrifuge tube.
  2. OPTIONAL: If desired, wash the tissue 3x with 5mL PBS to remove blood and other contaminants from the tissue.
  3. Add two times as much 3.2mm stainless steel beads (by mass) as you have sample (so for 1g sample, add 2g beads). One scoop ≈ 160mg.
  4. Add three volumes of buffer per mass of tissue (3mL of buffer per gram of tissue)
  5. Tightly screw the centrifuge tubes closed.
  6. Place tubes into the Bullet Blender™ 50.
  7. Set controls for SPEED 10 and TIME 9 minutes. Press start.
  8. After the run, remove the tubes from the instrument.
  9. Inspect samples. Fatty tissue homogenate will be difficult to see through due to the light scattering of lipid micelles formed, so it may be necessary to employ a pipette tip to check inside the tube for remaining pieces of intact tissue. If homogenization is unsatisfactory, run for another six minutes at SPEED 10.
  10. Proceed with your downstream application


  11. SAFETY NOTE!!!

    When using a centrifuge to separate your homogenate from the debris and beads, make sure your tubes are balanced.

    Typical Results
    Typical Results


The Bullet Blender™ home page shows the entire family of products and links to other applications.